Microwave frequency sweep interferometer for plasma density measurements in ECR ion sources: Design and preliminary results
Articolo
Data di Pubblicazione:
2016
Citazione:
Microwave frequency sweep interferometer for plasma density measurements in ECR ion sources: Design and preliminary results / Torrisi, G., Mascali, D., Neri, L., Leonardi, O., Sorbello, G., Celona, L., Castro, G., Agnello, R., Caruso, A., Passarello, S., Longhitano, A., Isernia, T., Gammino, S.. - In: REVIEW OF SCIENTIFIC INSTRUMENTS. - ISSN 0034-6748. - 87:2(2016). [10.1063/1.4933025]
Abstract:
The Electron Cyclotron Resonance Ion Sources (ECRISs) development is strictly related to the availability of new diagnostic tools, as the existing ones are not adequate to such compact machines and to their plasma characteristics. Microwave interferometry is a non-invasive method for plasma diagnostics and represents the best candidate for plasma density measurement in hostile environment. Interferometry in ECRISs is a challenging task mainly due to their compact size. The typical density of ECR plasmas is in the range 1011–1013 cm−3 and it needs a probing beam wavelength of the order of few centimetres, comparable to the chamber radius. The paper describes the design of a microwave interferometer developed at the LNS-INFN laboratories based on the so-called “frequency sweep” method to filter out the multipath contribution in the detected signals. The measurement technique and the preliminary results (calibration) obtained during the experimental tests will be presented.
Tipologia CRIS:
1.1 Articolo in rivista
Elenco autori:
Torrisi, G; Mascali, D; Neri, L; Leonardi, O; Sorbello, G; Celona, L; Castro, G; Agnello, R; Caruso, A; Passarello, S; Longhitano, A; Isernia, Tommaso; Gammino, S
Link alla scheda completa:
Pubblicato in: